BLACK-Comet Model Super Range 200-1100nm

Reference: STE-BLACK-Comet

StellarNet's concave grating spectrometer delivers high performance for spectroscopy applications in the UV-VIS wavelength ranges covering 190 - 850 nm, or optionally 280 - 900 nm.



0,00 € tax excl.




Eugenie Neureuther
Sales & Marketing
+49 8153 405-53
e.neureuther@laser2000.de
Product request: BLACK-Comet Model Super Range 200-1100nmX



Dear Frau Neureuther,

I am interested in the following product: BLACK-Comet Model Super Range 200-1100nm


I am interested in:

an offer
Information material
a call back and consulting
Please complete at least the mandatory fields marked with *.



* Title:

Academic Degree:

First Name:

* Last Name: * Last Name:

* Company Name: * Company Name:

Company Address:

* Street: * Street:

* Zip Code: * Zip Code:

* City: * City:


* E-Mail: * E-Mail:

* Phone: * Phone:

Text:




* What do you see in picture 2?* What do you see in Picture 2?

laserdiode laserschutzbrille laserschutzfenster leistungsmessgerät


Bitte auswählen:Please select:    Diode    Gauge    Safety Goggles    Safety Windows   



Please check red marked fields.




Your request has been sent successfully. You will be contacted as soon as possible.



EN.jpg


Product description

Concave Grating Spectrometer BLACK-Comet from StellarNet

StellarNet's concave grating spectrometer delivers high performance for spectroscopy applications in the UV-VIS wavelength ranges covering 190 - 850 nm (model C), or optionally 280 - 900 nm (model CXR). The instruments are exceptionally robust with no moving parts, and are packaged in small rugged metal enclosures (2.75 x 4 x 6 inch) for portable spectroradiometry and lab or process spectroscopy measurements. The BLACK-Comet UV-VIS concave grating spectrometer utilizes a 40 mm diameter concave grating with aberration correction to provide superb spectral imaging. This significantly improves spectral shapes by reducing comma and astigmatism found in plane grating spectrograph designs. The spectrograph architecture does not utilize mirrors, and therefore provides the lowest possible stray light in the UV with additional assistance from the holographic line grating.  The concave grating produces a flat field on the CCD detector creating uniform resolution over the entire range.

Advantages of the Concave Grating Spectrometer

One of the main advantages of the concave grating spectrometer BLACK-Comet is its ability to drastically reduce straylight levels, 0.02 % at 435 nm and 0.2 % at 200 nm, the lowest values of any field spectrometer. By design, the holographic grating has a smoother surface than the normal ruled gratings used in competing spectrometer models.  Surface roughness is the primary source of grating generated stray light and thus by employing superior grating technology, StellarNet’s Concave Grating Spectrometers are able to drastically reduce the effect. Additionally, grating scatter in the standard Czerny Turner design is collected and focused by the focusing mirror toward the exit port, this does not occur with the concave grating spectrometer  BLACK-Comet. When additional optical components like collimating and focusing mirrors are used in instrument design, it is more likely that light may be re-diffracted and scattered.  Similarly, due to the spectrometers ability to reduce light scatter, especially high power UV light, the Concave Grating spectrometer provides two to three times enhanced UV sensitivity as compared to plane grating/Czerny Turner optical configurations.

Sensitivity of the Concave Grating Spectrometer

Spectral sensitivity is one of the most important factors in choosing the proper miniature spectrometer. The concave grating optical design allows for high sensitivity and a smooth spectral distribution over the wavelength range as shown in the above spectrum of a Tungsten Halogen lamp.

Specifications of the Concave Grating Spectrometer

Dynamic Range:

2000:1 with 6 decades

Dimensions:

69 x 100 x 150 mm

Optical Resolution:

< 1 nm Resolving Res. with  25 µm slit

Power Consumption:

100 mA @ 5 VDC

Detector Type:

2048 pixel CCD, PDA opt.

Interface:

USB-2

Detector Range:

C = 190 - 850 nm or CXR = 280 - 900 nm

Data Transfer Speed:

3x / 40x faster than USB-1

Pixel Size:

14 um x 200 um

Detector Integration:

1 ms to 65 s

Concave Grating:

Aberration corrected

Slit Size Options:

14, 25, 50, 100, 200 µm

Grating Type:

Holographic, 590 g/mm

Stray Light:

0.02 % at 435 nm; 0.2 % at 200 nm

Spectrograph:

f/2, Flat field - No mirrors

Fiber Optic Input:

SMA905 0.22 na single fiber

Order Sorting Filter:

Integrated multi-band

Operating Systems:

Windows XP/Vista/7

Signal to Noise:

1000:1 CCD, PDA 2000:1

Software Included:

SpectraWiz program & apps

Digitizer:

16-bit

Also Free Programs For:

LabView / VC / VBA / Delphi